Using Oximeter Softkey Menus
Operator’s Manual
53
value for a particular patient in the limits screen. These values return to the factory or
institutional default values after power cycle.
Figure 36.
Setting Adult Limits Screen
To set limits
1. To adjust the upper and lower saturation and pulse rate limits, select the Adult or
Neonate Limit display. The Limit display includes the alarm limit table and current
measured SpO
2
and pulse rates. The title of the alarm limit table indicates whether the
oximeter is in Adult or Neonate monitoring mode. If SatSeconds or SPD are enabled, the
Limit display also includes the SatSeconds and SPD icons.
Note:
Limit changes remain in effect as long as the oximeter power continues, but return to the
institutional default limits when oximeter power is turned off. Only qualified service
personnel may change factory defaults to institutional defaults by following the procedures
outlined in the
N-600x Service Manual
.
2. Press the LIMITS softkey.
a. When Adult Limits are selected, the following current limits display.
Figure 37.
Adult and Pediatric Limits Screen
b. When Neonate Limits are selected, the following current limits display.
Figure 38.
Neonate Limits Screen
3. Press the ADULT or NEO softkey to select the Adult-Pediatric or Neonatal Limits screen.
4. Press the SELECT softkey to select the parameter to be adjusted.
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......