Study Results
Operator’s Manual
133
Study Results
Accuracy was calculated using the root mean square difference (RMSD).
Figure 103: Modified Bland-Altman Plot
Table 28.
SpO
2
Accuracy for Nellcor™ Sensors vs. CO-oximeters
SpO
2
Decade
MAX-A
MAX-N
MAX-FAST
Data
Points
Arms
Data
Points
Arms
Data
Points
Arms
60-70
71
3.05
71
2.89
71
2.22
70-80
55
2.35
55
2.32
55
1.28
80-90
48
1.84
48
1.73
48
1.48
90-100
117
1.23
117
1.68
117
0.98
1
Test Sensor;
Avg CO-oximeter value 70-100% SpO
2
2
Avg CO-oximeter value
70-100% SpO
2
Oximetry board with MAX-A sensor
Trendline of MAX-A sensor
Oximetry board with MAX-N sensor
Trendline of MAX-N sensor
Oximetry board with MAX-FAST sensor
Trendline of MAX-FAST sensor