![background image](http://html1.mh-extra.com/html/covidien/nellcor-oximax-n-600x/nellcor-oximax-n-600x_operators-manual_2673901114.webp)
Troubleshooting
108
Operator’s Manual
Note:
Once the oximeter is silenced, the oximeter sounds a piezo tone every three minutes as a
reminder of the primary speaker failure condition. The oximeter also sounds the piezo tone
to annunciate low, medium and high priority alarms during this time. Once an oximeter
reporting a primary speaker failure is powered off, it cannot be powered on again until
repaired.
To access primary speaker failure messages
The oximeter may detect a failure of the primary speaker and sound a high-pitched,
slow-pulsing piezo tone. A primary speaker failure message displays.
Figure 99.
Primary Speaker Failure Screen
1. Press HELP to continue. The following message displays.
Figure 100.
Speaker Failure Help Screen
2. Press BACK to display the speaker failure message again. The message cannot be cleared.
3. Press the ALARM SILENCE key to silence the slow-pulsing piezo tone.
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......