Troubleshooting
106
Operator’s Manual
READING TRENDS ...
None
Yes
Yes
When the
oximeter needs
more than
four to six
seconds to
retrieve trend
data from
memory.
Completely retrieve
sensor data or select
ABORT.
REPLACE SENSOR
None
No
Yes
When the sensor
or cable wiring is
defective.
Temporarily
acknowledge by pressing
the ALARM SILENCE key.
Replace sensor.
RESET DEFAULTS?
10
No
Yes
When attempting
to reset oximeter
to factory default
setting by
pressing the
RESET softkey on
the PARAMS
menu.
Leave default settings as
they are by selecting NO,
or revert back to the
original factory default
settings by selecting YES.
SAVE DEFAULTS?
10
No
Yes
When attempting
to save current
settings as power-
on defaults by
pressing SAVE in
the PARAMS
menu.
Confirm the new
institutional default
settings by selecting YES,
or leave default settings
as they are by selecting
NO.
SENSOR
DISCONNECTED
None
No
Yes
When the sensor
is disconnected
from the pulse
oximetry cable,
the cable is not
connected to the
oximeter, or the
cable wiring is
defective.
Temporarily
acknowledge by pressing
the ALARM SILENCE key.
Connect the cable and
sensor or the oximeter,
reconnect the sensor, or
check all connections. If
this does not clear the
condition, replace the
cable and/or sensor.
SENSOR TYPE
5
No
No
First message
displayed when a
sensor is
connected to the
oximeter.
Automatically clears on
time-out.
Table 16.
Common Prompts and Error Messages
Message
Time-out
in seconds
Exit on
Alarm
Exit on
Alarm
Silence
Displayed
Resolution
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......