Using Oximeter Softkey Menus
Operator’s Manual
47
To change the displayed language setting
a. With the oximeter in the normal monitoring mode, press the SETUP softkey.
b. Press the NEXT softkey.
c. Press the LANG softkey.
Figure 32.
Displayed Language Selection Screen
d. Use the ADJUST UP or ADJUST DOWN keys below the ALARM SILENCE key on the
oximeter panel to select the desired language.
e. Press the EXIT softkey.
Note:
The selected language displays until the oximeter is turned OFF. Any changes in language selection
do not last beyond power-off, but return to default English, unless qualified service personnel set the
selected language as a default by following the procedures outlined in the
N-600x Service Manual
.
5.
COMM Menu
—From the SETUP menu, press the NEXT softkey twice. Use the SELECT
softkey to toggle between the baud rate and communication protocol. Arrow up or down
to select the proper baud rate. Baud rates include 19200, 9600, and 2400 baud for all
protocols except SPDout, which allows 115200, 57600, and 19200 baud. Arrow up or
down to select the proper communication protocol. The OxiMax N-600x pulse oximeter
provides a bedside monitor communication for interfacing with the protocols listed
below. The factory default protocol is ASCII.
Table 6.
Communication Protocol Options
Communication
Protocols
N-600x using up to
Firmware 2.0
Default Baud Rate
N-600x with Firmware
2.0 and above
Default Baud Rate
ASCII
9600
9600
CLINICAL
19200
19200
GRAPH (Graphics)
9600
N/A
OXINET
9600
N/A
PHILIPS (VueLink)
19200
Use only 19200
SPACELBS (Spacelabs)
9600
N/A
MARQ (GE Marquette)
9600
N/A
DATEX (Datex-Ohmeda)
2400
N/A
SPDout
N/A
Default
115200
Available:
19200, 57600, 115200
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......