Data Port Connectivity
Operator’s Manual
79
Data Port Communications
To establish data port communication
1. Use the Serial Port Setup display to set the baud rate and the protocol of the data port on
the OxiMax N-600x pulse oximeter.
2. With the oximeter in the normal monitoring mode, press the SETUP softkey.
3. Press the NEXT softkey twice and then press the COMM softkey.
4. Press the SELECT softkey.
Figure 72.
Serial Port Setup Screen, Baud Rate Protocol Selection
5. Press the ADJUST UP or ADJUST DOWN keys to select the desired protocol. Protocol
options for the OxiMax N-600x pulse oximeter differ from those of the oximeter with the
SPD feature activated. The factory default protocol is ASCII. See
Table 10.
Communication Protocol Compatibility and Output
Communication
Protocols
N-600x with
Firmware up to 2.0
Default Baud Rate
N-600x with Firmware 2.0 and Above
Default Baud Rate
Output includes SPD
data fields
ASCII
9600
9600
Yes
1
CLINICAL
19200
19200
Yes
GRAPH (Graphics)
9600
N/A
No
OXINET
9600
N/A
No
PHILIPS (VueLink)
19200
19200
Use only 19200
No
SPACELBS (Spacelabs)
9600
N/A
No
MARQ (GE Marquette)
9600
N/A
No
DATEX
(Datex-Ohmeda)
2400
N/A
No
SPDout
N/A
Default
115200
Available:
19200, 57600, 115200
Yes
1
Not provided in real-time output
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......