Operating the Oximeter
44
Operator’s Manual
1.
MONITR Menu
—Isolate oxygenation (SpO
2
) or pulse (PULSE) trend data or view them
both together (DUAL) for a specified length of time. Establish trend data parameters for
specific time segments and set minimum and maximum trend values.
a.
VIEW Menu
—Isolate oxygenation (SpO
2
) or pulse (PULSE) trend data or view them
both together (DUAL). Access the HIST and Amplitude submenus by selecting the
NEXT Option.
-
DUAL
softkey
—View oxygenation (SpO
2
) and pulse (PULSE) trend data
together for a specified length of time.
-
SpO
2
softkey
—Isolate oxygenation (SpO
2
) trend data for a specified length of
time.
-
PULSE softkey
—Isolate pulse (PULSE) trend data for a specified length of time.
-
HIST softkey
—Choose to delete or keep current trend data history. To delete,
select YES. To keep, select NO and return to the HIST menu, go BACK, or EXIT to
the Main Menu.
-
AMP softkey
—Set the trend data view to “Pulse Amplitude Units” (PAU) using
the AMP (amplitude) Menu. This is an arbitrary unit of measure gauging the
distance from the peak to the valley of a pulse waveform. The trend data when
set to PAU can register amplitudes up to 200 PAU.
b.
ZOOM Menu
—Access to the TIME option, SCALE option, AUTO option, and BACK
option if the trend view is set in DUAL, SpO
2
, or PULSE mode.
Note:
The ZOOM menu will not allow access to the options listed if it is in AMP mode. SCALE and
AUTO options do not appear as options without trend data history to review.
-
TIME softkey
—Select from hours to minutes to seconds for viewing a specific
time segment of trend data. Cycle through the options by pressing the TIME
softkey. The options proceed in the following order: 48h, 36h, 24h, 12h, 8h, 4h,
2h, 1h, 30m, 15m, 40s, or 20s.
-
SCALE softkey
—Select the maximum and minimum values for the SpO
2
or
PULSE trend graph. Default SpO
2
trend values are 10 to 100. Default PULSE
trend values are 5 to 250. Cycle through the options by pressing the Select
softkey. The options proceed in the following order: ±5, ±10, ±15, ±20, ±25,
±30, ±35, ±40 and ±50 units variance.
-
AUTO softkey
—Obtain the maximum and minimum rounded values based on
all graphed trend data.
-
DELETE softkey
—Choose to delete or keep current trend data. To delete,
select YES. To keep, select NO and return to the previous menu.
2.
SENSOR Submenu
—The SENSOR menu is only available when using a single-
patient OxiMax adhesive pulse oximetry sensor. After connecting an OxiMax
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......