Help and Support
Operator’s Manual
101
4. From page (2 / 2) of the HELP MAIN window, press ADJUST DOWN to select PRDELAY and
then press SHOW. The HELP PRDELAY window appears.The SatSeconds help topic
contains a total of two (2) consecutive help windows. Press the NEXT softkey to scroll
through each window of the selected help topic for the following information:
“Pulse Rate Alarm Delay can reduce alarms reported for brief pulse rate limit violations.
The Pulse Rate Alarm Delay can be set to 5 or 10 seconds, or OFF.”
5. From page (2 / 2) of the HELP MAIN window, press ADJUST DOWN to select SATSECONDS
and then press SHOW. The HELP SATSECONDS window appears. The SatSeconds help
topic contains a total of six (6) consecutive help windows. Press the NEXT softkey to scroll
through each window of the selected help topic for the following information.
“SatSeconds can reduce alarms reported for mild or brief SpO
2
limit violations. Each SpO
2
violation can be described as a product of magnitude (number of percentage points of
the SpO
2
value falls outside the limit) and time (the number of seconds the SpO
2
the
value remains outside the limit). The SatSeconds limit sets the minimum value the
SatSeconds must reach before an alarm is reported. For example: if the SpO
2
lower alarm
limit is 90 and the measured SpO
2
value is 88, the resulting value is 2 after one second, 4
after two seconds, and so on. If the SatSeconds limit is set to 10, an alarm is reported after
five seconds. To adjust the SatSeconds limit, press LIMITS.”
6. Press the BACK softkey at any juncture to review the previous window(s). Continue to
press BACK to return to the HELP MAIN window.
7. Press EXIT to return to the oximeter’s Main menu.
To access a single help topic
Access single on-screen help topics by pressing the HELP/CONTRAST key from a
submenu. As an example, this procedure provides direction for accessing help
information on the SatSeconds feature.
1. Press LIMITS on the oximeter Main menu and then SELECT to highlight SAT-S
(SatSeconds).
2. Press the HELP/CONTRAST key. The HELP LIMITS window appears.
Figure 92.
Limits Help Screen Main Menu
3. Press ADJUST UP or ADJUST DOWN to highlight an available help topic (SELECT, NEO and
ADULT). For this example, highlight SELECT.
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......