Managing OxiMax™ Pulse Oximetry Sensor Data
Operator’s Manual
71
Managing the In-Sensor Data Type Feature
The In-Sensor Data Type display enables you to adjust the type of patient alarm
event trend data to be recorded in an OxiMax pulse oximetry sensor. OxiMax pulse
oximetry sensors can be set to record either “SpO
2
” or “SpO
2
+BPM.”
Note:
The OxiMax pulse oximetry sensor data type can only be set when the sensor is not
connected to the oximeter.
To select sensor data type
1. With the oximeter turned on and no cables attached to the SpO
2
OxiMax sensor port,
press the SETUP softkey.
2. Press the SENSOR softkey.
3. Press the DATA softkey.
Note:
OxiMax sensor data type settings are displayed on the oximeter screen. If no sensor is
connected, both sensor types and the full set of options for each are displayed. If a sensor is
connected, only the sensor data type for that sensor displays.
Figure 60.
Identifying In-Sensor Data Type Screen
4. Use the SELECT softkey to toggle between SENSOR-R (sensor contains read-only chip)
and SENSOR-RW (sensor contains a read/write chip).
Note:
The SENSOR-R feature supports all the current OxiMax pulse oximetry sensors. The
SENSOR-RW feature is only applicable to OxiMax pulse oximetry sensors with a read/write
chip installed.
5. Use the ADJUST UP or ADJUST DOWN key below the ALARM SILENCE key on the oximeter
to select the sensor data type. SENSOR-R and SENSOR-RW selections are:
•
SPO2
•
SPO2+BPM
•
DEFAULT
6. Press the EXIT softkey to set the sensor data type.
To view sensor event record
Access patient alarm event data with the patient sensor attached to the oximeter.
1. Press the TREND softkey from the main menu.
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......