Managing the View Display
Operator’s Manual
37
the SatSeconds circle icon and SPD triangle icon and their settings. The SPD alarm
sensitivity value appears just below the SPD icon. When the SPD feature is enabled, the
triangle icon fills from the bottom to the top as desaturation patterns develop. The
triangle icon empties from the top to the bottom as patterns dissipate. If the SPD icon
reaches full, it indicates a threshold violation and the SPD alarm sounds. The oximeter will
sound the SPD alarm sooner if the SPD alarm sensitivity is set to the default value of one
(1). A less sensitive setting will result in less frequent alarms. Visible when the SPD feature
is activated and the SPD feature is enabled. For more information about activation,
contact Nellcor’s Technical Services Department at 1.800.635.5267 or your local Nellcor
representative. For more information about using the feature, reference
When the oximeter is not attached to AC power, it runs on an internal battery. When
using the internal battery, a battery fuel gauge appears in the upper left corner of
the display.
To select the Real-Time Trend View
1. With the oximeter in the normal monitoring mode, press the SETUP softkey.
2. Press the VIEW, then NEXT softkeys.
3. Press the TREND softkey. The real-time trend view displays.
Figure 22.
Real-time Trend Display View
To set up the Trend Data Display
1. Press the SETUP softkey.
2. Press the VIEW, then NEXT softkeys.
3. Press the TREND softkey.
4. Press the VIEW softkey.
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......