Using Oximeter Softkey Menus
Operator’s Manual
49
Note:
When the oximeter is in the FAST response mode, the oximeter may produce more SpO
2
and pulse
rate alarms than expected. The response mode, however, may impact the SPD alarm behavior.
d. Use the ADJUST UP or ADJUST DOWN keys below the ALARM SILENCE key on the
oximeter panel to select the desired response mode.
e. Press the EXIT softkey.
LIGHT Menu
1.
OFF softkey
—Use the OFF softkey to turn the backlight off entirely. This option does not
remain after power cycle, but returns to the factory default brightness. Use ADJUST UP or
ADJUST DOWN keys below the ALARM SILENCE key on the oximeter panel to brighten or
dim the display backlight.
2.
EXIT softkey
—Exit to the main menu.
Adjusting the Factory Default Settings
Overview
WARNING
Audible alarms should not be silenced if patient safety could be compromised.
WARNING
Each time the pulse oximeter is used, check alarm limits to ensure that they are
appropriate for the patient being monitored.
WARNING
The SPD feature is for use with adults patients only. In Neonate Mode, the SPD feature
remains OFF.
Caution
Use of the SPD alarm feature does not change the need to set threshold limits appropriate to
the patient being monitored.
The OxiMax N-600x pulse oximeter is shipped with factory default settings. Factory
default settings are divided into two groups: adult and neonate. The oximeter lists
adult-pediatric alarm limit settings rather than neonate settings at power-up. Set the
oximeter’s operating mode to adult-pediatric or neonatal using the LIMITS softkey.
This setting remains active until the oximeter is turned OFF. Have a qualified service
technician set institutional default settings, if they are different than the power-on
default settings, using the
N-600x Service Manual
.
Use the softkeys to change alarm limits, displays, baud rates, time and date, and
trend data views. Some values cannot be saved as power-on default values.
•
The oximeter will not accept an %SpO
2
Lower Alarm Limit of less than “85” as a power-on default.
•
The oximeter will not accept an AUDIBLE ALARM OFF option as a power-on default.
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......