Maintenance
9-10
Operator's Manual
and an alarm message and help notes are displayed (see section
provides SSI ranges and a list of the possible causes and action to correct for each, if
necessary.
Table 9-5.
Signal Strength Indicator Troubleshooting Chart
SSI indication
Description/cause
Action
5 bars
STATUS CODE: 4
Full strength signal
No action required.
1 through 4 bars—stable
Signal is weaker, but is strong and stable
enough to generate an accurate rSO
2
value
No action required.
1 through 4 bars—fluctuating
and unstable
Signal is weak and unstable
Signal interference
Identify any possible sources of interfer-
ence, such as electrocautery or similar elec-
trical equipment interferences, intense
light or radiation sources, moving artifacts
or perturbations on the mains.
Sensor application over nevi, sinus cavities,
the superior sagittal sinus, subdural or epi-
dural hematomas or other anomalies such
as arteriovenous malformations, broken
skin, thick fatty deposits, hair or bony pro-
tuberances, or area contaminated with
residue
Clean the area, reapply the sensor to the
patient in an adjacent area; follow the
sensor application instructions and make
sure no hair, anomalies, or residues are
under the sensor.
Defective sensor
Apply a new sensor to the patient.
Moisture in sensor connection
Dry the cable and sensor connectors. Make
sure the connection is not exposed to
moisture.
0 bar
See error messages in data channel area on
screen (see
)
Follow on-screen instructions.
Summary of Contents for INVOS 5100C
Page 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Page 12: ...Page Left Intentionally Blank x Operator s Manual...
Page 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Page 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Page 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Page 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Page 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Page 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Page 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Page 177: ......