Status codes are numbers associated with specific conditions that are displayed in the Tabular
trends screen as lines of data in Run function and Case archive function (see section
,
). The status codes help the user to identify
alarm and status message conditions and troubleshoot the system setup.
Alarms and Status Messages with rSO
2
Values Displayed
During normal monitoring situations when rSO
2
values are being displayed, alarm and status
messages may appear at the top of the Main screen. Actions in
are listed to assist the
operator to try to correct a problem, if necessary. Additional operating indications to assist the
user will appear on the screen at the right of the corresponding affected channel label indicator
(for example, two-digit rSO
2
values turn yellow when below baseline desaturation threshold).
Table 6-5.
Status Messages—rSO
2
Value Displayed
Status message
Status code
Color, location
Cause
Action
ALARM HIGH accompa-
nied by alarm tone
5
Red, top right; corre-
sponding rSO
2
value is
red
Upper alarm limit is
exceeded.
Profile patient.
ALARM LOW accompa-
nied by alarm tone
6
Red, top right; corre-
sponding rSO
2
value is
red
Lower alarm limit is
exceeded.
Profile patient.
BATTERY LOW accompa-
nied by alarm tone
9
Yellow, top right
Battery power is critical.
Plug in A/C power and
turn master (mains)
switch on.
COM PORT UNAVAIL-
ABLE
12
Yellow, top right
Output device not con-
nected.
Check connections with
Com port device (PC).
PRINTER UNAVAILABLE
13
Yellow, top right
Printer not connected.
Check connections with
Com port device (PC),
power connection, and
online status.
USB FULL
20
Yellow, top right
USB flash drive memory
is full.
Insert a blank flash drive.
USB STORAGE ERROR
14
Yellow, top right
USB flash drive error.
Check USB flash drive
connection. Insert a
flash drive.
DATE AND TIME
N/A
Green, top left
N/A
N/A
ADULT, PEDIATRIC OR
NEONATAL
N/A
Green, top center
Adult, pediatric, or
infant/neonatal applica-
tion dependent on
sensor model used.
N/A
Summary of Contents for INVOS 5100C
Page 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Page 12: ...Page Left Intentionally Blank x Operator s Manual...
Page 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Page 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Page 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Page 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Page 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Page 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Page 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Page 177: ......