Run Function (Patient Monitoring) in Normal Mode
Operator's Manual
6-7
or baselines may be set for all channels in use at once, individually for a selected channel or chan-
nels and manually to a previous rSO
2
.
Set Current rSO
2
as Baseline for All Channels
To set a baseline, press the HOME
key to display the Main screen, then perform the following
navigation bar options:
The baseline value or values will be set at the current rSO
2
value or values and the menu will return
to the Main screen. This will display an event mark line and code on the screen in the graph of
each channel. The baseline value or values will not be saved if the INVOS™ system is turned off.
Note:
If a baseline cannot be set preinduction, it is recommended that a desaturation threshold be set using the
manual baseline feature.
Set Individual Channels to Current rSO
2
Baseline
Set Individual Channels to Current rSO
2
Baseline—Two-Channel Monitoring
To set an individual channel baseline, press the HOME
key to display the Main screen, then
perform the following navigation bar options:
Select the channel or channels to set the baseline at the current rSO
2
value. This will display an
event mark line and code on the screen in the graph area for the individual channel only.
When finished, press the HOME
key to exit the Baseline menu and return to the Main screen.
Set Individual Channels to Current rSO
2
Baseline—Four-Channel Monitoring
BASELINE MENU
EVENT MARK
ALARM AUDIO ON/OFF
NEXT MENU
SET BASELINES
SET CHANNEL
MANUAL SET
PREVIOUS MENU
BASELINE MENU
EVENT MARK
ALARM AUDIO ON/OFF
NEXT MENU
SET BASELINES
SET CHANNEL
MANUAL SET
PREVIOUS MENU
CHANNEL 1 SET
CHANNEL 2 SET
PREVIOUS MENU
MAIN MENU
Summary of Contents for INVOS 5100C
Page 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Page 12: ...Page Left Intentionally Blank x Operator s Manual...
Page 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Page 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Page 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Page 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Page 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Page 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Page 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Page 177: ......