4-1
4 Quick Setup
4.1
Overview
This chapter briefly describes the setup of the INVOS™ regional saturation patient monitoring
system (see also Chapter
for com-
plete instructions).
4.2
Initial Setup
1.
Plug in the INVOS™ system monitor to a power source.
Note:
The blue
LED illuminates to indicate the power is connected and the battery is charging. If the blue
light is not on, turn the back panel master (mains) power switch on.
2.
Connect the preamplifier or preamplifiers to the monitor. Align the red dot on the preamplifier cable
silver connector with the red dot on side panel input connection. Insert the cable connector into the
side panel connection while keeping the dots aligned. Be sure to fully insert the connector until it
locks.
Figure 4-1.
Connecting the Preamplifier to the Monitor
3.
Connect the reusable sensor cable connectors to the preamplifier or preamplifiers. Use the color
coding.
Summary of Contents for INVOS 5100C
Page 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Page 12: ...Page Left Intentionally Blank x Operator s Manual...
Page 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Page 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Page 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Page 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Page 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Page 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Page 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Page 177: ......