Battery Recharge
Operator's Manual
9-3
9.5.1
Check Battery Status
1.
Place the monitor on a sturdy, flat surface and plug it in to a power source (mains) by attaching the
power cord to the inlet (see
).
Note:
Use only hospital-grade AC mains outlets.
2.
Turn the mains power switch on. The blue
LED on the key panel illuminates to indicate the power
is connected and the battery is charging (see
3.
Access Run function by pressing the ON/OFF
key to display the Start screen, then select NEW
PATIENT or PREVIOUS PATIENT from the navigation bar options.
4.
Turn the mains power switch off. A battery symbol/gauge
should appear in the lower right
side of the display screen, indicating the battery status. If the symbol shows a full battery, there is no
need for battery charging. If the symbol shows a partial or total discharge of the battery, turn the mains
power switch on and the battery recharges automatically.
9.5.2
Recharge the Battery
1.
Place the monitor on a sturdy, flat surface and plug it in to a power source (mains) by attaching the
power cord to inlet (see
Note:
Use only hospital-grade AC mains outlets.
2.
Turn the mains power switch on. The blue
LED on the key panel illuminates to indicate the power
is connected and the battery is charging (see
). The monitor can be on or off. The battery
recharges as long as the blue LED is on.
Summary of Contents for INVOS 5100C
Page 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Page 12: ...Page Left Intentionally Blank x Operator s Manual...
Page 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Page 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Page 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Page 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Page 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Page 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Page 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Page 177: ......