8-1
8 Database Operational Mode and Area Under the
Curve
8.1
Overview
Effective January 1, 2008, the Society of Thoracic Surgeons (STS) has added cerebral oximetry as
a new metric to be collected in its adult cardiac surgery database. This chapter describes the
metrics, as well as how to capture and download the metric data with the INVOS™ regional sat-
uration patient monitoring system. Importantly, this will allow all users access to area under the
curve calculations.
WARNING:
For area under the curve (AUC) to calculate you must set left and right baselines prior to
induction in real-time or accept the auto-baseline value when prompted by the device
(approximately 5 minutes after monitoring begins). Baselines should not be reentered at any
other time or the integrity of AUC will be compromised. Similarly, entering skin closure will cease
calculation of AUC.
Note:
Database software is not compatible with cases collected using previous software versions. Therefore,
previous cases stored on the INVOS™ system should be downloaded and saved elsewhere prior to
loading the database software.
8.1.1
INVOS™ System Database Application Software
Summary of Contents for INVOS 5100C
Page 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Page 12: ...Page Left Intentionally Blank x Operator s Manual...
Page 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Page 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Page 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Page 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Page 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Page 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Page 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Page 177: ......