ILLUSTRATED PARTS BREAKDOWN
Syringe Module
7A-13
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
Figure 7A-6.
Top Disk Holder - Front Case Assembly
1
Install
4½”
Gasket (item
281
) into channel on Front
Case.
1
351
281
Top Disk Holder Assembly
(see
CAUTION
statement)
NOTE:
All parts illustrated in this figure are provided
separately and as assembled items in the Front
Case Assembly Kit (Table
9
).
312
2
(see
CAUTION
statement)
315
311
2
2
Spring (item
312
) torque is set by inserting end of
spring in hole of Flag Disk Holder (item
311
) and
winding holder counterclockwise.
The Top Disk Holder illustrated here is not
backward compatible with older versions of the
Front Case Assembly. Prior to replacing this part,
reference the “Corrective Maintenance” chapter
and Figure 7A-7 for replacement instructions and
part identification.
CAUTION
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......