5.2.6
Removing IUI Board Assembly
1. If present, remove Retainer Clip from
Dual Connector Cable-IUI Board
connection.
2. Disconnect Dual Connector Cable from
IUI
Board
J2
.
3. Disconnect Syringe Size Sensor
harness from
IUI
Board
J4
.
4. Disconnect Flex Cable from
IUI
Board
J3
.
5. Use small diagonal cutters to lift and
remove snap rivet from
IUI
Board.
6. Remove
IUI
Board Assembly.
During Reassembly:
Route Dual Connector Cable and Syringe
Size Sensor harness under
IUI
Support
Bracket.
CORRECTIVE MAINTENANCE
Syringe Module
5A-8
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Dual Connector Cable
IUI Board Assembly
Snap Rivet
Flex Cable
Syringe Size
Sensor
Harness
IUI Support Bracket
Do not cut rivet.
CAUTION
Retainer Clip