CHECKOUT AND CONFIGURATION
2-6
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
3. To change
KVO
rate, press
Change
Rate
soft key and enter new rate (
0.01
to
2.5
mL/h in
0.01
mL/h increments).
NOTE:
Flow rates less than
0.01
mL/h
are available only with a
1
cc syringe.
For larger syringes, the lower limit
adjusts to
0.1
mL/h.
4. To change
KVO
volume, press
Change
Volume
soft key and enter new volume
level (
0.5
to
5%
).
5. To accept changes(s), press
CONFIRM
soft key.
• Settings change and
System Config
- Syringe
page
1
displays.
2.4.6
Max rate
The
Maximum rate
value identifies the
maximum rate at which a bolus dose can be
delivered. When the
Max Rate
option is
selected during bolus dose programming,
the bolus is delivered at the maximum rate
for an automatically calculated duration.
1. After accessing
System Config -
Syringe
options display, page
2
, press
Max rate
soft key.
2. To change maximum rate, press
Change Value
soft key and enter new
rate (
0.1
to
99.9
mL/h in
0.1
mL/h
increments;
100
to
999
mL/h in
1
mL/h
increments).
2.4
CONFIGURATION SETUP - SYRINGE
MODULE
(Continued)
2.4.5
KVO
(Continued)
S
y
stem Configuration - S
y
ringe
Change
Value
Maximum rate
adjustment
Maximum
rate:
mL/h
_ _ _
S
y
stem Configuration - S
y
ringe
Change
Value
Confirm
Maximum rate
adjustment
Maximum
rate:
mL/h
999
S
y
stem Configuration - S
y
ringe
KVO
Enable
Disable
Change
Rate
Change
Volume
Rate:
2
mL/h
Volume:
_5%
Allows use of the KVO feature.
CONFIRM
S
y
stem Configuration - S
y
ringe
KVO
Enable
Disable
Change
Rate
Change
Volume
Rate:
_ 2
mL/h
Volume:
2%
Allows use of the KVO feature.
CONFIRM
-- Continued Next Page --
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......