NOTES:
(Continued)
• Pressing the
CONFIRM
soft key while in a
System Configuration option screen:
♦
accepts existing setting or setting
change
♦
displays next option setting screen (if
applicable) or returns display to
System
Config - Syringe
screen
• Pressing the PC Unit
CANCEL
key while in
a System Configuration option screen:
♦
leaves setting unchanged
♦
returns display to
System Config -
Syringe
screen
To access System Configuration options:
1. Hold
OPTIONS
key at power up.
2. Press
PAGE DOWN
soft key.
3. Press
Syringe Module
soft key.
4. To view additional options, press
PAGE
DOWN
soft key.
2.4.1
All Mode
Enabling
All mode
allows
ALL
to be
selected as the volume to be infused (
VTBI
),
to deliver the entire contents of the syringe.
CHECKOUT AND CONFIGURATION
2-3
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
>Select an Option
or EXIT
EXIT
Factory default:
No
System Config - Module 1 of 2
PC Unit
Shared Infusion Settings
Pump Module
SPO2 Module
PAGE
DOWN
2.4
CONFIGURATION SETUP - SYRINGE
MODULE
(Continued)
>Select an Option
or EXIT
EXIT
Syringe Module
System Config - Module 2 of 2
PAGE
UP
EtCO2 Module
>Select an Option
or EXIT
EXIT
S
y
stem Config - S
y
ringe 2 of 2
Pressure limit-Disc:1000
mmHg
PAGE
UP
Pressure limit-NoDisc: Med
Near End:
Priming:
Disabled
Max rate:
999
mL/h
Disabled
>Select an Option
or EXIT
EXIT
S
y
stem Config - S
y
ringe 1 of 2
PAGE
DOWN
Back Off:
Disabled
Disabled
Fast Start:
KVO:
Auto Pressure:
Disabled
Disabled
Disabled
All Mode:
If
KVO
and
ALL
modes are both enabled,
the
ALL
mode will not deliver the entire
syringe contents. A small percentage of the
syringe contents will be left available for
KVO
operation.
CAUTION
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......