5A-13
CORRECTIVE MAINTENANCE
Syringe Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.11 Removing Pressure Sensor Board
Assembly
1. Disconnect Pressure Sensor Harness
from Pressure Sensor Board.
2. Remove four (
4
) screws attaching Base
Disk Holder to Top Disk Holder and
Front Case.
3. Remove Base Disk Holder and Pressure
Sensor Board Assembly.
During Reassembly:
When installing an old version of Base Disk
Holder, use
4-40 x
5
/
16
screws (item
486
) in
upper mountings, and
4-40 x
3
/
16
screws
(item
484
) in lower mountings. When
installing a newer version of Base Disk
Holder, use
4-40 x
5
/
16
(item
486
) for all
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Pressure Sensor Board Assembly
Pressure Sensor
Harness
Base Disk Holder (old)
To avoid damage to the Front Case when
reinstalling the Base Disk Holder, use
correct size screws (reference Figure
7A-4
).
CAUTION
Base Disk Holder (new)
(See CAUTION statement)
No boss
Boss
4-40 x
5
/
16
(4 PL)
4-40 x
5
/
16
(upper mountings)
4-40 x
3
/
16
(lower mountings)
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......