5B-19
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.15 Removing Syringe Barrel Clamp
Assembly
(Continued)
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Syringe Barrel
Clamp Assembly
Shaft Seal
(see installation illustration)
Compression
Spring
Shaft
Support
Bushing
Potentiometer
Extension Guide
Park Position Collar
and Setscrew
Retaining
Rings
Shaft Seal Installation
Front Case
Recess
Bushing
Shaft Seal
Front Case
Shaft Seal
(fully seated)
Press around edges
until seated in recess.
Step 1
Step 2