5B-12
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.9
Removing Internal Frame Assembly
1. Remove one (
1
) screw attaching three
(
3
) ground wires to
IUI
Support Bracket.
2. Remove three (
3
) screws attaching
Internal Frame to Front Case Assembly.
3. Remove
IUI
Bracket Strap.
4. Remove one (
1
) screw attaching
IUI
Support Bracket to
IUI
Bracket.
5. Remove
IUI
Support Bracket and
IUI
Bracket.
6. Remove two (
2
) screws attaching
Internal Frame to Drivetrain Assembly.
7. Remove Internal Frame.
During Reassembly:
• Route Linear Sensor Harness between
IUI
Bracket and Internal Frame.
• Route Lock Assembly harness through
top of and under Internal Frame so that
Frame does not pinch harness when
attached to Front Case, and harness can
reach Logic Board
J13
.
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Drivetrain
Assembly
IUI Bracket
Ground wires
IUI Bracket Strap
IUI Support Bracket
Front Case
Assembly
Linear Sensor Harness
Internal Frame
Lock Assembly
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......