5B-22
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
5.2.18 Removing Latch Lever Assembly
1. Remove two (
2
) screws attaching Latch
Cover to inside of Front Case.
2. Remove Ground Harness.
3. Remove Latch Lever and Latch Seal.
During Reassembly:
1. Install Latch Seal onto shaft of Latch
Lever.
2. Install Latch Lever into Front Case,
using a blunt instrument to seat seal into
oval recess.
3. Attach Latch Cover to Front Case.
Latch
Cover
Latch
Lever
Latch
Seal
Latch Seal
installs into
oval recess