CORRECTIVE MAINTENANCE
Syringe Module
5A-14
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.12 Removing Top Disk Holder
1. Remove two (
2
) screws attaching Top
Disk Holder to Front Case.
2. Remove Top Disk Holder.
NOTE:
When replacing the Top Disk Holder,
remove the Flag Disk Holder and Spring from
the old part and reuse.
During Reassembly:
• Set spring torque by inserting end of
Spring in hole of Flag Disk Holder and
winding holder counterclockwise.
• Ensure Silicone Tubing and Gasket in
Top Disk Holder mounting area on Front
Case are in place and undamaged. (See
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Flag Disk Holder
Spring
Top Disk Holder
To replace “Older Version” Top Disk Holder
(see below), replace with Front Case
9
).
CAUTION
“Older Version”
Top Disk Holder
See
CAUTION
statement above.