5B-4
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
5.2.2
Removing IUI Connector Assemblies
Remove two (
2
) screws attaching each
IUI
(left and right) to module.
During Reassembly:
• Ensure ground clips are still installed on
both
IUI
connectors.
• To install Right
IUI
Connector Gasket,
remove protective backing and adhere to
IUI
Connector.
• To install Left
IUI
Connector Seal,
position seal on one end of connector
and stretch to other end to conform to
connector body. Gently press on seal to
seat completely. Use lint-free swab to
apply alcohol to top, sides, and bottom of
seal for lubrication while installing to Rear
Case. Do not apply alcohol to contacts
or connector.
IUI Connector, Left
IUI Connector, Right
IUI Connector Seal, Left
IUI Connector Gasket, Right
Ground clip at each mounting
hole on both IUI Connectors.
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......