5B-11
CORRECTIVE MAINTENANCE
PCA Module
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
Internal Frame Assembly
5.2.8
Removing Logic Board Assembly
1. If present, remove Retainer Clip from
Dual Connector Cable-Logic Board
connection.
2. Disconnect Dual Connector Cable from
Logic Board
J8
.
3. Disconnect Linear Sensor Harness from
Logic Board
J12
.
4. Remove one (
1
) screw attaching Logic
Board to
IUI
Support Bracket.
5. Remove one (
1
) screw attaching Logic
Board to Internal Frame Assembly.
6. Disconnect harnesses from following
locations on Logic Board:
J4
(display)
J7
(motor)
J11
(flange detect)
J13
(lock harness)
7. Remove Logic Board.
During Reassembly:
Route Dual Connector Cable and Linear
Sensor Harness under
IUI
Support Bracket.
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Dual Connector Cable
Logic Board
J12
Linear Sensor Harness
IUI Support Bracket
J8
J11
J4
J7
J13
Retainer Clip
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......