3. To accept change, press
Confirm
soft
key.
• Setting changes and
System Config
- Syringe
page
2
displays.
2.4.7
Near End
Enabling
Near-End-of-Infusion
(
NEOI
)
allows an alert to be set from
1
to
60
minutes before the infusion is complete.
The alert occurs at the configured time or
when
25%
of the
VTBI
remains, whichever
occurs later.
NOTE:
In the following procedure, the
Near
End
option is changed to
Enable
. Changing
the option setting to
Disable
disables the
Change Time
soft key, as displayed in
step
1
.
1. After accessing
System Config -
Syringe
options display, page
2
, press
Near End
soft key.
2. To change option setting, press soft key
next to applicable option (
Enable
or
Disable
).
3. To change
Alert Time
, press
Change
Time
soft key and enter new time
(
1
to
60
minutes).
4. To accept changes(s), press
CONFIRM
soft key.
• Settings change and
System Config
- Syringe
page
2
displays.
2.4
CONFIGURATION SETUP - SYRINGE
MODULE
(Continued)
2.4.6
Max rate
(Continued)
CHECKOUT AND CONFIGURATION
2-7
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
Disable
Disable
CONFIRM
S
y
stem Configuration - S
y
ringe
Change
Time
Disable
Enable
Near-End-of
Infusion
Alert Time:
2 min
Allows use of the Near-End-of
Infusion feature.
Disable
Disable
CONFIRM
S
y
stem Configuration - S
y
ringe
Change
Time
Disable
Near-End-of
Infusion
Alert Time:
2 min
Allows use of the Near-End-of
Infusion feature.
Enable
Disable
Disable
CONFIRM
S
y
stem Configuration - S
y
ringe
Disable
Near-End-of
Infusion
Alert Time:
10 min
Allows use of the Near-End-of
Infusion feature.
Enable
Change
Time
S
y
stem Configuration - S
y
ringe
Maximum rate
adjustment
Maximum
rate:
mL/h
99.9
Change
Value
Confirm
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......