On the Trig I/O A and Trig I/O B tabs there are the following control parameters:
Input pulse trigger type
– Selection of incoming trigger type. (None, Stimulus
trigger, Video trigger, Mute trigger).
Minimum input pulse length (µs)
- Minimum length of accepted trigger pulse
Output pulse length (µs)
– Length of output pulse from trig out.
Enabled
– If set incoming trigger activates trigger out pulse. If input trigger type
in
Stimulus
or
Video
, output trigger will be sent immediately when input trigger is
detected by the Main Unit. In case trigger type is Mute, Output trigger will be
sent when muting is complete set on.
On the Non-isolated 8-bit trigger input tab there are the following control
parameters:
Enabled
– If set allows incoming triggers to non-isolated 8-bit trigger input.
Minimum input pulse length (µs)
- Minimum length of accepted trigger pulse.
On the Amplifier Muting tab there are the following control parameters:
Magnetic stimulus input mute (Main amplifier)
– If set allows the main
amplifier is muted, when mute trigger comes in.
Magnetic stimulus input mute (Switch)
– If set allows the switches are muted,
when mute trigger comes in.
Mute start difference
- Time difference (in µs) between switch mute and main
amplifier mute. If value is positive switches are muted first.