Bipolar Connection Box interface:
1.
Input connectors
– 5-pole Binder connector.
2.
Reference inputs
–
Black safety connectors.
3.
Active inputs
–
Red safety connectors.
4.
Ground (GND)
–
Electric ground, used for modules/sensors that are not directly
electrically connected to skin like GSR module, breathing belt, piezo sensors etc.
5.
Bipolar Out
–
26 pin KEL-connector for connection to Tesla amplifier.
6.
Power ON (yellow) LED
–
Indicates that Bipolar is connected to Tesla amplifier
and power is turned on.
For each input there are two types of connectors: safety connectors (active and reference)
and 5-pole Binder connector.
Default labeling for input are “33 - 40”. GND input is safety
connector type.
As default Bipolar Connection Box
is configured for “Bipolar ExG” use (EMG, EEG and
ECG). Then each input has additional 9.1 kOhm protective resistor, no signal attenuation.
All possible configurations (using internal DIP switches) for Bipolar Connection Box are
explained in Appendix 9.
SEE APPENDIX 9 BEFORE CHANGING CONFIGURATION IN
BIPOLAR CONNECTION BOX!
4.8 Video Option
The NeurOne video option consists of:
Video camera (Full HD, audio input, HDMI output)