Page 138 of 191
800581-2.11 NeurOne System User Manual.doc
8
8
S
S
P
P
E
E
C
C
I
I
F
F
I
I
C
C
A
A
T
T
I
I
O
O
N
N
S
S
NeurOne system
Intended use:
Electroencephalograph
8.1 Main unit (model Extended)
Intended use:
Electroencephalograph
Storage methods:
Online
to PC (Raw)
Amplifiers (EXG) :
Max. 4 pcs 40 channel amplifier boxes ("EXG" )
or
Amplifiers (Tesla):
Max. 4 pcs 40 channel amplifier boxes ("Tesla" )
Power source:
Powerbox, AC power adapter, 100-240 V AC / 50
–
60Hz, Max 1.7 A (12 V DC ---- 5A) (UL60601-1
compatibility)
or
PROTEC POWER, Model: PMP60-12
Power Input voltage:
12 V
Current consumption:
Max 3.0 A
Connectors:
Analog Out (16 ch, Centronics 24 pin)
Audio Out (3.5mm connector for headphones/external
audio amplifier)
Isolated trigger A Out/In
Isolated trigger B Out/In
8-bit trigger IN (8 input lines, non-isolated, D-Sub 15)
Isolated SPI socket for external sensors
Ethernet connection to PC
Optical link for fibres to Amplifier (4 pcs)
Isolated Amplifier power connector (10 V output, 4 pcs)
Serial port (3 pcs)
Enclosure class:
IP20
Display:
Two row display, 24 character/row
Loudspeaker:
Integrated
Dimensions:
33 x 12 x 27 cm
Weight:
3.6 kg
8.2 Main unit (model Black)
Storage methods:
Online
to PC (Raw)
Amplifiers (EXG) :
Max. 4 pcs 40 channel amplifier boxes ("EXG ")
or
Amplifiers (Tesla):
Max. 4 pcs 40 channel amplifier boxes ("Tesla" )