![BIO RAD CFX384 Instruction Manual Download Page 46](http://html1.mh-extra.com/html/bio-rad/cfx384/cfx384_instruction-manual_2753185046.webp)
CFX96 and CFX384 Systems Manual
37
•
Increment.
Enter a temperature to increment the selected step; the increment amount is
added to the target temperature with each cycle
•
Ramp Rate.
Enter a rate for the selected step; the range depends on the block size
•
Time.
Enter a hold time for the selected step
•
Extend.
Enter a time to extend the selected step. The extend amount is added to the
hold time with each cycle
•
Beep.
Check the box to include a beep at the end of the step
TIP: When you enter a number that is outside the option range, the software
changes the number to the closest entry within the range.
Delete Step Button
To delete a step in the protocol:
1. Select a step in the graphic or text view.
2. Click the
Delete Step
button to delete the selected step.
WARNING!
You cannot undo this function.
Temperature Control Mode
The instrument uses one of two temperature control modes to determine when the sample
reaches the target temperature in a protocol.
TIP: The sample volume can be changed before a run by editing the Sample
Volume parameter in the Start Run tab (see “Start Run Tab” on page 24).
Enter a sample volume in the protocol editor to select a temperature control mode:
•
Calculated mode.
When you enter a sample volume between 1 and 50 μl (96-well block)
or between 1 and 30 μl (384-well block) the thermal cycler calculates the sample
temperature based on the sample volume. This is the standard mode
•
Block mode.
When you enter a sample volume of zero (0) μl, the thermal cycler records
the sample temperature as the same as the measured block temperature
Protocol AutoWriter
Open the Protocol AutoWriter to quickly write protocols for PCR and real-time PCR
experiments. To open the Protocol AutoWriter, select one of these options:
•
Click the
Protocol AutoWriter
button in the main software window toolbar
•
Select
Tools > Protocol AutoWriter
from the menu bar in the main software window
Summary of Contents for CFX384
Page 9: ...Table of Contents viii ...
Page 49: ...Protocols 40 ...
Page 89: ...Data Analysis Overview 80 ...
Page 149: ...Resources 140 ...
Page 157: ...Index 148 ...