CFX96 and CFX384 Systems Manual
135
WARNING!
Never heat the block after adding a cleaning solution. Heating the
block with cleaning solution will damage the block, reaction module, and thermal
cycler base.
•
Clean the inner lid.
Use a soft, lint-free cloth and water to remove debris and solutions
from the inner lid surface. Never use abrasive detergents or rough material that will
scratch the surface. Cleaning the inner lid improves precise sample heating and cooling.
Application Log
Before starting a new run, an instrument initiates a self-diagnostic test to verify that it is
running within specifications. The software records results of this test in the Run log and
Application log file. If you notice a problem in one or more experiments, open the run and
application logs to find out when the problem started happening.
CFX Manager software tracks information about the state of an instrument during a run in the
Application Log
(Figure 122). Use these logs to track events that occur on instruments and in
the software and for troubleshooting.
To open the Application log in the main software window, select
View > Application Log
.
Figure 122. Example of an Event Log file.
Software Help Tools
This CFX Manager software provides Help with the following tools:
•
Select the
Search
or
Index
tabs in this Help site to search for more information
•
Open the
Glossary
to look up words that are specifically used in this software. For
widely used words, consult a PCR dictionary or glossary
•
Press the
F1
key on your keyboard to open software help about topics in many of the
software windows
•
Print any Help page by right-clicking on it and selecting
•
Click the
www.bio-rad.com
link at the bottom of each Help page to open the Bio-
Rad website for links to information and resources
Summary of Contents for CFX384
Page 9: ...Table of Contents viii ...
Page 49: ...Protocols 40 ...
Page 89: ...Data Analysis Overview 80 ...
Page 149: ...Resources 140 ...