Omnisense 7000S/8000S
15-13
Practice Mode
When using the practice mode, the application operates in the
same manner as in the regular measurement mode described in
Chapter 8, "Patient SOS Measurement‚" page 8-1, except for
the following differences:
•
After the user carries out the second measurement, the
application calculates and provides the Coefficient of Vari-
ance (CV) value and measurement results by means of
a
score,
according to a CV- score (precision-score) table (see
below). From that point on, following each new measure-
ment, the application provides the recalculated CV score.
•
The CV score is given per measurement site, and the CV
score is site dependent, accordingly.
CV score is a function of the Standard Deviation and your aver-
age measurements values. Your CV score should fall in the
range stated in the column labeled “Good”, in Table 14.1, for
each skeletal site before you start measuring patients.
Table 15.1 CV scores
•
The user can print the practice session results or data, but
these cannot be saved. If the
Save
button is pressed, a pop-
up dialog box opens, reading: “The Practice results/data
cannot be saved”.
CV Score//Site
Good
Needs Improvement
Poor
Distal 1/3 Radius
≤
0.40
>
0.40,
≤
0.80
>0.80
Proximal Phalanx
III
≤
0.81
>
0.81,
≤
1.62
>1.62
Metatarsal V
≤
0.66
>
0.66,
≤
1.32
>1.32
Summary of Contents for Sunlight Omnisense 7000S
Page 4: ...iv ...
Page 12: ...xii ...
Page 18: ...xviii ...
Page 22: ...xxii ...
Page 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Page 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Page 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Page 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Page 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Page 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Page 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...