12-2
User Guide
CHAPTER 12
Cleaning and Disinfecting Omnisense
2.
Use any cleaning detergent or mild soap suitable for external
cleaning of electrical equipment to clean the Omnisense
device. Follow the instructions of the detergent or soap man-
ufacturer.
P
RECAUTIONARY
N
OTES
:
•
Do not allow any cleaning liquid to get into the probe con-
nector sockets, the disk drives or the connector sockets and
openings on the rear panel.
•
Do not use cleaning agents that can damage or destroy plas-
tics, such as ammonia, acetone, hydrochloride acid (HCL),
etc.
The Ultrasonic Probe
CLEANING
The patient contact surface of the probe should be cleaned
immediately after use. For cleaning, it is not necessary to dis-
connect the probe from the Omnisense unit.
The following steps should be followed for cleaning the probe:
1.
If a probe cap was used, remove and dispose it before clean-
ing.
2.
To clean the probe of any acoustic gel or residues, use only
dry, soft tissues or towels. If required, use towels that are
pre-moistened with a mild detergent or soap
.
3.
The probe should be allowed to air dry if pre-moistened tow-
els are used for cleaning.
4.
Visually inspect the probe to ensure cleanliness.
P
RECAUTIONARY
N
OTES
:
•
Do not immerse the probe in water, liquid detergents or
soaps.
•
DO NOT USE scrubbing materials or hard brushes that
might scratch the probe head surface.
Summary of Contents for Sunlight Omnisense 7000S
Page 4: ...iv ...
Page 12: ...xii ...
Page 18: ...xviii ...
Page 22: ...xxii ...
Page 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Page 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Page 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Page 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Page 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Page 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Page 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...