Measuring SOS at the Radius
actions. When the probe is in proper contact with
the area to be scanned, an audible signal is heard
from Omnisense.
Move the probe about 70 degrees first to the lateral
side, and then the same amount to the medial side,
so that the probe executes an arc of about 140
degrees. One full probe movement, i.e. a distance of
140 degrees in one direction, should take 4-5 sec-
onds.
Figure 8.10 Scanning the radius
To insure accurate and reproducible results, note the
following:
•
The edge of the probe toward the palm side should follow
the marked line on the skin at all times.
Correct scan procedure for the radius.
Top row photos first, from left to right:
1. Scan begins with probe vertical to the
positioned arm.
2. Scan by moving first to the lateral
direction, about 70 degrees.
3. Continue the scan by returning the
probe to the starting point.
4. Then scan in the opposite (medial)
direction, about 70 degrees, and finally
return again to start point. Repeat until
measurement cycle is complete.
2
1
3
4
Summary of Contents for Sunlight Omnisense 7000S
Page 4: ...iv ...
Page 12: ...xii ...
Page 18: ...xviii ...
Page 22: ...xxii ...
Page 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Page 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Page 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Page 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Page 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Page 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Page 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...