8-20
User Guide
CHAPTER 8
Patient SOS Measurement
•
Never hold the probe at an angle to the surface of the limb;
maintain the surface of the probe perpendicular to the bone
at all times.
•
Do not hold the probe with only one hand.
•
Do not use a circular motion. The probe movement should
follow parallel to the marked line.
•
The probe must be
moved
across the full range of 140
degrees, around the axis of the bone, while maintaining a
position perpendicular to the bone. Do not merely
tilt
the
probe around the marked line on the skin.
•
Check to insure that a scan movement through 140 degrees
takes about 4-5 seconds. The scan is being conducted too
slowly or too quickly if the movement takes longer or
shorter, respectively, than this amount of time.
•
Maintain the probe’s surface in contact with the patient’s
skin throughout the scan. No part of the probe surface
should be raised from the skin at any point during the scan.
•
During the measurement cycle, concentrate on the site and
correct application of the probe. Avoid looking at the
screen. Use the audible signal to provide feedback and
guide you in correct execution of the scan.
•
If probe caps are used, replace the probe cap with every
measurement.
If the measurement is not going well, you can
restart at any time during the cycle by pressing
Restart
(for the first cycle the button is labeled
Start
).
The foot pedal is provided to allow hands-free
operation of Omnisense during a measurement
cycle. Pressing on the foot pedal is equivalent to
clicking on the
Start
/
Restart
button.
If Omnisense is not receiving an appropriate signal
— indicated by an intermittent sound or no sound
— check that sufficient ultrasound gel is on the
probe as well as on the area being scanned.
Summary of Contents for Sunlight Omnisense 7000S
Page 4: ...iv ...
Page 12: ...xii ...
Page 18: ...xviii ...
Page 22: ...xxii ...
Page 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Page 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Page 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Page 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Page 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Page 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Page 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...