8-6
User Guide
CHAPTER 8
Patient SOS Measurement
statistically meaningful conclusions from an SOS result,
the conditions of the measurement should be as close as
possible to those used in building the baseline data.
Therefore,
the exact procedure given in this Guide must
be followed for each site
.
Overall, the method of measurement with Omnisense is similar
regardless of the site being measured, although there are some
important points that vary from site to site. In this chapter the
correct method for measuring the distal 1/3 radius is presented
in detail. A careful reading of the two procedures which follow
will ensure that you are familiar with the key points of correct
measurement and will minimize inaccuracies in your results.
When measuring at other skeletal sites, remember to check
carefully for variations in the method. See Chapter 15, "Site
Measurement Methodologies‚" page 15-1.
Positioning the limb and finding the site
S
TEP
1 Position the patient and the limb
SOS measurement at the distal 1/3 radius has been
defined by Sunlight clinical research at a precise
point on the limb which facilitates the easiest and
most reproducible measurement. This point is half-
way between the elbow and the end of the middle
finger. However, to insure compatibility with the
reference database, it is important to locate the site
precisely.
Always measure the patient’s non-dominant side. In case
of a disorder or impairment in the non-dominant limb,
such as edema, a mass, undetachable ring or IV line, use
the contra-lateral side instead.
Seat the patient in a chair, opposite the operator and
close to a standard desk.
Summary of Contents for Sunlight Omnisense 7000S
Page 4: ...iv ...
Page 12: ...xii ...
Page 18: ...xviii ...
Page 22: ...xxii ...
Page 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Page 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Page 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Page 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Page 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Page 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Page 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...