2. General
2.3
ICMP HDD status LEDs
ICMP HDD status LEDs
5
4
Image 2-3
Status overview PWR/ERROR and READY LEDs:
4
5
HDD I/O
RAID
ICMP HDD Status
Off
Off
HDD idle / Disk in RAID OK.
Blinking
Green
Off
HDD I/O activity / Disk in RAID OK.
Blinking
Green
Blinking Red
HDD I/O activity / RAID rebuilding.
Off
Red
HDD idle / Disk error. See troubleshooting table below for curative
actions.
Troubleshooting
Situation
Solution
One disk failed (red LED) +
RAID degraded
.
The ongoing event is not interrupted.
Note:
The disk status (RAID degraded) can be
retrieved via the (Web) Commander. See user
guide of the (Web) Commander.
1. Switch off the power.
2. Replace the defect HDD with an original HDD
spare part. See procedure "Removing a HDD
from the ICMP", page 71, and "Installing a
HDD into the ICMP", page 72. Ensure to insert
the HDD
fi
rmly.
3. Switch on the power.
Result:
As soon the new HDD is detected by the
ICMP the rebuild of the RAID is started (Blinking
red LED).
One disk failed (red LED) + Error 10580 “
local
storage not available
”.
Note:
The disk status (Error code) can be
retrieved via the (Web) Commander. See user
guide of the (Web) Commander.
1. Switch off the power.
2. Replace the defect HDD with an original HDD
spare part. See procedure "Removing a HDD
from the ICMP", page 71, and "Installing a
HDD into the ICMP", page 72. Ensure to insert
the HDD
fi
rmly.
3. Switch on the power.
Result:
As soon the new HDD is detected by the
ICMP the rebuild of the RAID is started (Blinking
red LED).
R5906594 ICMP 09/11/2016
9
Summary of Contents for ICMP
Page 1: ...ICMP Service Manual DP Alchemy R5906594 05 09 11 2016 ...
Page 4: ......
Page 6: ...Table of contents 2 R5906594 ICMP 09 11 2016 ...
Page 18: ...2 General 14 R5906594 ICMP 09 11 2016 ...
Page 20: ...3 Preventative maintenance actions 16 R5906594 ICMP 09 11 2016 ...
Page 70: ...4 Troubleshooting 66 R5906594 ICMP 09 11 2016 ...
Page 88: ...5 ICMP service procedure 84 R5906594 ICMP 09 11 2016 ...