C.3 Clearing an S2 warning
After the firmware update, the AT600 may show an S2 warning. If so, complete the following steps:
1.
In the update software, click on the com port button and select the specific port that was used for the update
(see top red box in figure 39 below).
2.
Click on the clear S2 warning button (see bottom red box in figure 39 below).
Figure 39: Clearing an S2 warning
3.
After about 15 seconds, a screen similar to figure 40 below is displayed. Click OK and then reboot the AT600 to
confirm that the S2 error has been cleared.
Figure 40: S2 warning successfully cleared
C.4 Support
If the firmware update was unsuccessful, reboot the AT600 and repeat the procedure described in this appendix. If
there is still a problem, send an email to
and describe the problem in
detail.