Chapter 6. Communication
108
AquaTrans™ AT600 User’s manual
Command 187 (0xBB): Write alarm configuration
This command is to write the alarm configuration.
Table 89: Request data bytes for write alarm configuration
Table 90: Response data bytes for write alarm configurations
2
Error
Invalid selection
3 - 4
Undefined
5
Error
Too few data bytes received
6
Error
Device-specific command error
7
Error
In write protect mode
8 - 15
Undefined
16
Error
Access restricted
8 - 127
Undefined
Byte
Format
Description
0
Unsigned-8
Channel number (1/2)
1
Unsigned-8
Measurement type
2
Unsigned-8
Alarm value unit
3 - 6
Float
Alarm value
7
Unsigned-8
Alarm type:
0: Low;
1: High;
2: Fault
8
Unsigned-8
Alarm state:
0: Normally;
1: Failsafe;
Byte
Format
Description
0
Unsigned-8
Channel number(1/2)
1
Unsigned-8
Measurement type
2
Unsigned-8
Alarm value unit
3 - 6
Float
Alarm value
7
Unsigned-8
Alarm type:
0: Low;
1: High;
2: Fault
8
Unsigned-8
Alarm state:
0: Normally;
1: Failsafe;
Code Class
Description
Summary of Contents for Panametrics AquaTrans AT600
Page 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Page 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Page 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Page 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Page 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Page 173: ......