AquaTrans™ AT600 User’s manual
99
Chapter 6. Communication
Command 171 (0xAB): Set loop current percentage
This command is to set the output percentage of loop current.
Table 59: Request data bytes for set loop current percentage
Table 60: Response data bytes for set loop current percentage
Table 61: Command-specific response codes for set loop current percentage
5
Error
Too few data bytes received
6
Error
Device-specific command error
7
Error
In write protect mode
8
Undefined
9
Error
Incorrect loop current mode or value
10 - 15
Undefined
16
Error
Access restricted
17 - 31
Undefined
32
Error
Busy
33 - 127
Undefined
Byte
Format
Description
0 - 3
Float
Loop current percentage, units of percent.
Byte
Format
Description
0 - 3
Float
Loop current percentage, units of percent.
Code Class
Description
0
Success
No command-specific errors
1 - 2
Undefined
3
Error
Passed parameter too large
4
Error
Passed parameter too small
5
Error
Too few data bytes received
6
Error
Device-specific command error
7
Error
In write protect mode
8
Undefined
9
Error
Incorrect loop current mode or value
10 - 15
Undefined
16
Error
Access restricted
17 - 31
Undefined
32
Error
Busy
33 - 127
Undefined
Code Class
Description
Summary of Contents for Panametrics AquaTrans AT600
Page 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Page 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Page 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Page 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Page 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Page 173: ......