Chapter 4. Initial setup and programming
50
AquaTrans™ AT600 User’s manual
4.6.2.4
Setting the alarm
Setting the measurement type
Press [√] to open a screen similar to the
following.
Use the [
] or [
] key to select a specific
digit, then use the [
] or [
] key to change
that digit as required. After all digits are
correct, press [x] to save. Then, verify at your
frequency counter that you see the
frequency you entered. If desired, you can
repeat this procedure with several different
frequencies. After testing is complete, press
[x] to return to the digital output menu.
The alarm can be either an open circuit (normally closed type)
or a short circuit (normally open type), depending on the error
condition. To check the alarm or change its settings, in the
digital output menu select alarm and press [√]. The following
screen opens.
Select measurement and press
[√]. A screen similar to the
following opens.
In the measurement menu,
select the type of analog output
to be used, and press [√]. The
screen returns to the previous
display.
Summary of Contents for Panametrics AquaTrans AT600
Page 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Page 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Page 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Page 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Page 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Page 173: ......