Appendix A. Specifications
144
AquaTrans™ AT600 User’s Manual
A.4 General
A.4.1 Wiring cable specifications and requirements
•
Cable diameter range for PWR connection: 7 to 12mm, refer to gland Hole 1 on Figure 24 on page 20
•
Cable diameter range for Hart, Modbus and I/O connection: 5 to 8mm, refer to gland Hole 2,3 and 4 in Figure 24 on
•
Temperature range of cable for PWR, Hart, Modbus and IO connection: 14° to 185°F (-10° to 85°C)
The cable should meet the CE and UL standard below:
• Conductor cross section solid range: 0.2 mm2 to 2.5 mm2
• Conductor cross section stranded range: 0.2 mm2to 2.5 mm2
• Conductor cross section stranded, with ferrule without plastic sleeve range: 0.25 mm2 to 1 mm2
• Conductor cross section stranded, with ferrule with plastic sleeve range: 0.25 mm2 to 1 mm
• Conductor cross section AWG/kcmil range: 12 to 26 AWG according to UL/CUL range: 14 to 28
A.4.2 Cable fixing requirement and gland torque
Refer to Figure 24 on page 20 for the gland hole position.
To make a reliable IP67 sealing performance of the enclosure during cabling, the gland must be tightened well, below
torque value is a reference to make a reliable NEMA 4X/IP67 sealing between cable and gland:
•
Operation torque for gland hole 1 and 5: 2.7 N.M
•
Operation torque for gland hole 2, 3 and 4: 2.5 N.M
A.4.3 Display languages
English/Chinese/German/French/Italian/Japanese/Portuguese/Russian/Spanish/Swedish/Turkish
Note:
The meter will be set to the language specified by the customer before shipping.
A.4.4 Product models
Based on the line power type, the AT600 ultrasonic flow meter is available in two series:
•
Models of AC meter: 85-264VAC, 50-60Hz, 10W, up I AT6-**-****-****-*-1-*-**-**-*-*, AT6KIT-*1, AT6KIT-*2, AT6KIT-*3
and AT6KIT-*7
•
Models of DC meter: 12-28VDC, 10W, up I AT6-**-****-****-*-2-*-**-**-*-*, AT6KIT-*4, AT6KIT-*5, AT6KIT-*6 and
AT6KIT-*8
Note:
* in the product model name is either a number from 0-9 or a letter from A-Z.
Summary of Contents for Panametrics AquaTrans AT600
Page 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Page 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Page 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Page 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Page 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Page 173: ......