AquaTrans™ AT600 User’s manual
103
Chapter 6. Communication
Table 73: Command-specific response codes for read pulse configuration
Command 178 (0xB2): Read frequency configuration
This command is to read the frequency configuration.
Table 74: Request data bytes for read frequency configuration
Table 75: Response data bytes for read frequency configuration
Table 76: Command-specific response codes for read frequency configuration
Code Class
Description
0
Success
No command-specific errors
1
Undefined
2
Error
Invalid selection
3 - 4
Undefined
5
Error
Too few data bytes received
6
Error
Device-specific command error
7 - 127
Undefined
Byte
Format
Description
0
Unsigned-8
Channel number (1/2)
Byte
Format
Description
0
Unsigned-8
Channel number
1
Unsigned-8
Measurement type
2
Unsigned-8
Frequency value unit
3 - 6
Float
Frequency base value
7 - 10
Float
Frequency full value
11 - 14 Unsigned-32
Full frequency, unit is Hz
15
Unsigned-8
Frequency error handling:
0: Low;
1: High;
2: Hold;
3: Value
16 - 19 Unsigned-32
Error handling value, unit is Hz
Code Class
Description
0
Success
No command-specific errors
1
Undefined
2
Error
Invalid selection
3 - 4
Undefined
5
Error
Too few data bytes received
6
Error
Device-specific command error
7 - 127
Undefined
Summary of Contents for Panametrics AquaTrans AT600
Page 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Page 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Page 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Page 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Page 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Page 173: ......