AquaTrans™ AT600 User’s manual
119
Chapter 6. Communication
Table 129: Response data bytes for write transducer information
Table 130: Command-specific response codes for write transducer information
Command 205 (0xCD): Write transducer traverses and spacing
This command is to write transducer traverses and spacing.
Table 131: Request data bytes for write transducer traverses and spacing
0 - 3
Unsigned-32
Transducer type:
0: Other;
4 - 7
Unsigned-32
Transducer frequency
8 - 11
Unsigned-32
Transducer wedge type
12 – 15 Unsigned-32
Transducer wedge angle
16 – 19 Unsigned-32
Transducer wedge SOS
20 - 23 Unsigned-32
Transducer Tw
Byte
Format
Description
0 - 3
Unsigned-32
Transducer type:
0: Other;
4 - 7
Unsigned-32
Transducer frequency
8 - 11
Unsigned-32
Transducer wedge type
12 – 15 Unsigned-32
Transducer wedge angle
16 – 19 Unsigned-32
Transducer wedge SOS
20 - 23 Unsigned-32
Transducer Tw
Code Class
Description
0
Success
No command-specific errors
1 - 4
Undefined
5
Error
Too few data bytes received
6
Error
Device-specific command error
7
Error
In write protect mode
8 - 15
Undefined
16
Error
Access restricted
17 - 127
Undefined
Byte
Format
Description
0
Unsigned-8
Transducer traverse
1 - 4
Float
Transducer spacing
Summary of Contents for Panametrics AquaTrans AT600
Page 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Page 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Page 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Page 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Page 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Page 173: ......