Figure 4-34
4.4.9 Error Condition
The error annunciator on the front panel display tells about the error
condition of the multimeter. If there are one or more syntax or
hardware errors found, the error annunciator will be lit to inform you.
The multimeter stores errors using the first-in-first-out (FIFO) order
and it records up to 20 errors in the error queue. The first error
detected will be the first error retrieved. Refer to Chapter 6 for the list
of error messages.
5.
The error annunciator will be turned off when you have read
the errors in the memory.
6.
If there is no error found when you check the error
queue, the multimeter shows “
NO ERRORS
”.
7.
If there are more than 20 errors found, the last error in
the error queue will be replaced with “-350”, means too many
errors.
8.
The error queue is cleared when the multimeter is turned
off, or when a *CLS command is received.
112