How to specify a delay time
You can set the delay time from either the front panel operations or
the remote interface operations.
Front Panel Operation
To set the delay time, press MENU button and use
◁
and
▷
buttons to
locate “TRIG” option. Press ENTER to select it. Then again find “DELAY”
option using
◁
and
▷
buttons. Press ENTER to select it. Then use
◁
and
▷
buttons to move through the digits and
△
and
▽
buttons to set
the desired delay time (between 0 to 3600 second). The locations of
Measurement Function
Setting
Trigger Delay
Time
DCV/DCI
PLC >= 1
1.5 ms
PLC < 1
1.0 ms
Ω2 and Ω4
(PLC >= 1)
100Ω ~ 100kΩ
1.5 ms
1 MΩ
15 ms
10 MΩ ~ 100 MΩ
100 ms
Ω2 and Ω4
(PLC < 1)
100Ω ~ 100kΩ
1.0 ms
1 MΩ
10 ms
10 MΩ ~ 100 MΩ
100 ms
ACV/ACI
(Remote Interface/
External Trigger/
Single Trigger)
3 Hz
7.0 s
20 Hz
1.0 s
200 Hz
600 ms
ACV/ACI
(Front Panel w/ Auto
Trigger On)
3 Hz
1.5 s
20 Hz
200 ms
200 Hz
100 ms
Frequency/Period
Remote Interface /
External
1.0 s
Front Panel w/ Auto
Trigger ON
0 s
79