Speed of Scanner Card Measurement
AutoZero OFF,AutoGain OFF,AutoRange OFF, Scan Timer=0,60Hz
single
function(VDC)
NPLC
Take Time with 2000 Readings(sec)
rate(ch/s)
(Fast 4.5) 0.02
68
29.4
(Slow 4.5 & Fast 5.5) 0.1
74
27.0
(Slow 5.5 & Fast 6.5) 1
105
19.0
(Slow 6.5) 10
408
4.9
AutoZero OFF,AutoGain OFF,AutoRange OFF, Scan Timer=0,60Hz
Mix function
(VDC+2WRES)
NPLC
Take Time with 2000 Readings(sec)
rate(ch/s)
(Fast 4.5) 0.02
306
6.5
(Slow 4.5 & Fast 5.5) 0.1
318
6.3
(Slow 5.5 & Fast 6.5) 1
442
4.5
(Slow 6.5) 10
1710
1.2
How to program each channel with measurement function
Press SHIFT + DIGITS for scanning configuration. Use
◁
and
▷
to scroll through submenus. Press ENTER on “SET SCAN CHA”
submenu. Use
◁
and
▷
to switch from channel 1 to channel 10, and
use
△
and
▽
to switch among the measurement functions. Leave an
unused channel with a blank space (----). Press ENTER when ten
channels are all defined.
How to configure scan count, scan interval and store
measurement readings
Press SHIFT + DIGITS for scanning configuration. Use
◁
and
▷
buttons to scroll through submenus. Press ENTER to select
“OPERATION” submenu. Use
◁
and
▷
to switch among scan count
107